HI-Volume Air Flow Calibrator,HFC-Series
The HFC-Series,HI-Volume Air Flow Calibrators,have eliminated the need for cumbersome orifice plates and water manometers.The HFC-XXC series units utilize a precision machined Venturi tube coupled with a pressure differential gauge giving a direct reading in the volumetric units of your choice (i.e.SCFM @stp).The unit is calibrated against an in-line N.I.S.T.traceable laminar flow element.The primary calibrator meets the requirements of MIL Std.45662A.HFC-XX series units are intended to be used open to air.
The direct meter read-out will indicate the flow in standard CFM,LPM,or CMM at standard conditions (29.92"of Hg and 70 º F)).Given actual sampling tempera- ture &barometric pressure during calibration,a technician can convert actual flow readings (i.e.ACFM)to standard units (i.e.SCFM)by making a simple calculation using look-up correction factors from tables given in the operating manual.
Depending on orientation at calibration (vertical vs.horizontal),HI-Q offers two HFC models:The HFC-XX is intended for high flow calibrations in the verticalplane,such as seen with the “HVP ” style high volume air samplers..Whereas HI-Q ’s HFC-SIDE-XX is intended for the calibration of low to high flow rates in the horizontal plane,such as seen with HI-Q ’s line of portable continuous duty and grab air samplers.
The HFC-XX series units have a standard 1 ½"male straight pipe thread (MSPT)which can be screw into most low to medium volume air sampling pumps (ie:HI-Q ’s,CF-1001 or CF-901 ).On higher volumetric flow samplers,a user must incorporate an adapter plate.With the use of HI-Q ’s FHA-4CF 4"diameter adapter disc,an operator is able to calibrate a unit with a 4"diameter filter holder assembly.The FHA- 810CF adapter plate is used to calibrate units with an 8"x 10"paper holder.
Ordering Information Replace “XX ” with the uppermost desired flow rate..If a dual scale reading is required (i.e.both CFM &LPM displayed on one scale)replace “C ” with “DSC ” ((e.g.order:HFC-50DSC for 10-50 CFM &280-1500 LPM marked scale) High Volume,Single Range Air Flow Calibrators ranges be- tween 10-50 CFM (280-1500 LPM)